60 Operating Guide
6 Configuration Tasks
To Use DHCP to Provide the GC IP Address
To set the GC to use DHCP:
1 Turn off the GC.
2 While pressing and holding [Prep Run] and [Stop] on the
GC keypad, turn on the GC. This will set the GC to use
DHCP to obtain an IP address.
3 Ping the GC using the static IP address or hostname
assigned to the GC by your network administrator. See
the Troubleshooting manual for details or if the GC does
not respond.
You can also set the GC to use DHCP using the software
keypad.
1 Click [Options]. Scroll to Communications and click [Enter].
2 Set Enable DHCP to On (press [On/Yes]).
3 Reboot the GC.