8-144
Performance Tests
8648D Test Record
NOTE If the automated Power Level Accuracy performance test is performed, the
RF Level Accuracy performance test is not required.
Table 8-88. RF Level Accuracy Performance Test - Part 2
Frequency
(MHz)
Amplitude
(dBm)
Limits (dBm)
Uncertainty
(±dB)Lower Measured Upper
2000 13 12 14 0.15
10 9 11 0.15
43 50.15
5.9 6.9 4.9 0.15
15.9 16.9 14.9 0.12
2499 13 12 14 0.32
10 9 11 0.32
43 50.32
5.9 6.9 40.32
15.9 16.9 14.9 0.25
3200 13 11.5 14.5 0.46
10 8.5 11.5 0.46
4 2.5 5.5 0.46
5.9 −7.4 4.4 0.46
15.9 17.4 14.4 0.41
4000 13 11.0 15.0 0.46
10 8.0 12 0.53
4 2.0 6.0 0.53
5.9 7.9 3.9 0.53
15.9 17.9 13.9 0.48