EasyManuals Logo
Home>Agilent Technologies>Test Equipment>8647A

Agilent Technologies 8647A User Manual

Agilent Technologies 8647A
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #472 background imageLoading...
Page #472 background image
9-6
Supplemental Verification Tests
CW Frequency Accuracy Supplemental Verification Test
Table 9-5. 8648A Option 1E5 Frequency Accuracy Supplemental Verification Test
Frequency
(MHz)
Limits (Hz)
Uncertainty
(±Hz)Lower Measured Upper
0.5 1.0 1.0 1.0
1 1.0 1.0 1.0
10 2.0 2.0 1.0
50 8.0 8.0 1.0
100 15.0 15.0 1.0
500 75.0 75.0 1.0
1000 150 150 1.0
Table 9-6. 8648B Option 1E5 Frequency Accuracy Supplemental Verification Test
Frequency
(MHz)
Limits (Hz)
Uncertainty
(±Hz)Lower Measured Upper
0.5 1.0 1.0 1.0
1 1.0 1.0 1.0
10 2.0 2.0 1.0
50 8.0 8.0 1.0
100 15.0 15.0 1.0
500 75.0 75.0 1.0
1000 150 150 1.0
2000 300 300 1.0