Specifications
Agilent Cary 3500 UV-Vis Spectrophotometer User’s Guide 15
For models that have Peltier temperature control, clear and unobstructed airflow
of the inlet and outlet must be maintained. Ambient air must be allowed to enter
the inlet and other heat sources such as exhausts, computers or other nearby
Peltier modules must be avoided.
For models without Peltier control, you must allow for the circulation of air for
cooling. Allow for 10 cm (4 inches) of space on both sides and approximately
10 cm (4 inches) in the rear for electrical connections and air flow.
Electrical specifications
For detailed power requirements see the Site Preparation Checklist.
The power cord and all other connections are located at the rear of the instrument
or on the side of the module. The power switch is located on the rear of the engine
and on the side panel of the module. Position the instrument for easy access to
the disconnecting device.
Mains Supply
Power cords are provided based on the user’s country requirements. Only the
supplied power cord is to be used with this equipment.
The installation of electrical power supplies must comply with the rules and/or
regulations imposed by local authorities responsible for the supply of electrical
energy to the workplace.
Shock Hazard
Danger of electrocution. Good electrical grounding is essential to avoid
potentially serious shock hazards. A 3-wire outlet with ground connection
must be provided for the Cary 3500 UV-Vis. Make certain that power outlets
are earth-grounded at the grounding pin.
For safety reasons, a separate power outlet receptacle should be provided for
each unit in the system. The use of extension cords or outlet adaptors is not
recommended.
If necessary, replace the power cord only with a cord equivalent to the one
specified in the Site Preparation Checklist.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Cary 3500 and is the answer not in the manual?

Agilent Technologies Cary 3500 Specifications

General IconGeneral
BrandAgilent Technologies
ModelCary 3500
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals