Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 43
Introduction to Your System 2
General Purpose Agilent ChemStation Software for UV-Visible Spectroscopy — Overview
Processing with Standards
In the fourth task, standards are used in addition to your sample data. This
requires an extension to the above concept to handle standards. Two
independent sets of containers are implemented, one for standards, and
another for samples. All processing containers are also duplicated. As with the
sample-only processing, all evaluation steps are done in parallel on both the
samples and the standards.
The evaluation in quantification is now a calibration using standards. The
coefficients are calculated based on the method’s settings and the current
standards in the Agilent ChemStation memory. This means that results now
become a function of a measured set of standards and of the standard
concentrations specified.
Raw Data
Processed
Spectra
Used
Wavelength
Results
Spectral Processing
Data Access
Data Evaluation
Raw Data
Processed
Spectra
Used
Wavelength
Calibration
Results
Calibration
Samples Standards