Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 55
Good Measurement Practices 4
General Considerations
Apertured Cells
In applications where sample volume is limited, apertured or microcells are
used. The width of these cells is reduced to reduce the volume and the blank
part of the cell must be blackened to avoid unwanted transmission and
reflection through the side walls. If the side walls are not blackened the result
will be poor photometric accuracy and, if different concentrations are
measured, poor linearity.
The disadvantage of apertured and microcells is that part of the light beam is
blocked. Not all the light passes through the sample and there can be some loss
of sensitivity. See Figure 5 for recommended and Figure 6 for cells you should
not use with the instrument.
Figure 5 Recommended cells
Quartz cells Quartz cells with black
CAUTION
* Quartz cells with black apertures smaller than 2 mm, when used with a multicell
transport, can lead to measurements with poor reproducibility.

Table of Contents

Other manuals for Agilent Technologies Cary 8454

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Cary 8454 and is the answer not in the manual?

Agilent Technologies Cary 8454 Specifications

General IconGeneral
BrandAgilent Technologies
ModelCary 8454
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals