Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 103
Using your UV-Visible Spectroscopy System 5
Quantitative Analysis using a Calibration with Standards
Analysis
1 Remember the orientation of the cell in the cell holder. Lift the level to
release the cell, remove it and flush it three times with about 1 ml of your
caffeine sample (1:1 dilution with distilled water). Then fill the cell with
about 3 ml of the caffeine sample. Make sure that the cell windows are clean
and reposition the cell in the same orientation as for the standard
measurement. Close the cell holder’s level.
2 Press the Sample button on the front of your spectrophotometer or click
Sample on the Instrument Panel to start the measurement.
3 Enter your sample’s information in the Sample Information dialog box and
click OK. The view switches to the samples and your quantitative results
will be displayed with the Sample/Result table.

Table of Contents

Other manuals for Agilent Technologies Cary 8454

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Cary 8454 and is the answer not in the manual?

Agilent Technologies Cary 8454 Specifications

General IconGeneral
BrandAgilent Technologies
ModelCary 8454
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals