EasyManuals Logo

Agilent Technologies Cary 8454 User Manual

Agilent Technologies Cary 8454
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #73 background imageLoading...
Page #73 background image
Agilent Cary 8454 UV-Visible Spectroscopy System Operator’s Manual 73
Using your UV-Visible Spectroscopy System 5
Measuring Caffeine Absorbance at 273 nm
6 Press the Blank button on the front of the spectrophotometer or click Blank
on the Instrument Panel to start the measurement.
7 Remember the orientation of the cell in the cell holder. Lift the level to
release the cell, remove it and flush it three times with about 1 ml of your
caffeine sample. Then fill the cell with about 3 ml of the caffeine sample.
Make sure that the cell windows are clean and reposition the cell in the
same orientation as for the reference measurement. Close the cell holder’s
level.
8 Press the Sample button on the front of your spectrophotometer, or click
Sample on the Instrument Panel to start the measurement.
9 The view shows you the spectrum of your caffeine sample with a vertical
line indicating your wavelength of interest. Below the spectrum is the
Sample/Result Table which shows the absorbance reading at 273 nm.

Table of Contents

Other manuals for Agilent Technologies Cary 8454

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Cary 8454 and is the answer not in the manual?

Agilent Technologies Cary 8454 Specifications

General IconGeneral
BrandAgilent Technologies
ModelCary 8454
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals