Power Meter Operation
Self Test
2-76 Agilent E4419B User’s Guide
Remote Testing
To invoke the remote self test, the IEEE 488.1 compliant standard 
command, *TST? is used. This command runs a full self test and returns 
one of the following codes:
• 0 - no tests failed
• 1 - one or more tests failed
The remote self test consists of the following tests:
• ROM checksum
• RAM 
• RAM Battery
• Display Assembly
• Calibrator
• Measurement Assemblies
• Communications Assembly (Implicit)
The communications assembly is tested implicitly, in that the command 
will not be accepted or return a result unless the GP-IB interface is 
functioning correctly. 
Refer to “Test Descriptions”, on page 2-77 if you require a description of 
each individual test.
When the *TST? command is executed, the screen is cleared. As each test 
takes place, the name of the test is listed on the screen. While a test is 
running, the message “Testing...” appears next to the name of the test. 
As each stage of the test is completed, the message “Testing...” is 
replaced by either the message “Passed” or “Failed”. 