Power Meter Operation
Setting the Resolution
2-38 Agilent E4419B Users Guide
Setting the Resolution
The resolution of each of the power meters windows can be set to four
different levels (1, 2, 3 or 4).
These four levels represent:
1, 0.1, 0.01, 0.001 dB respectively if the measurement suffix is
dBm or dB.
1, 2, 3 or 4 significant digits respectively if the measurement suffix
is W or %.
The default value is 0.01 dB (3 digits).
To set the resolution on the currently selected window:
1. Press . The current setting of the resolution is highlighted
on the softkey.
2. To change this setting press until the
required resolution setting is highlighted.
Meas
Setup
Resolution 1 2 3 4
Resolution 1 2 3 4
HP4402.book Page 38 Thursday, November 30, 2000 1:39 PM