Error Messages
Introduction
4-2 Agilent E4419B Users Guide
Introduction
This chapter contains information about the power meters error
messages. It explains how to read the power meters error queue and lists
all of the power meters error messages and their probable causes.
When there is a hardware related problem, for example, a power sensor
overload, the error message is displayed on the status line at the top of the
display. In addition, the errors are also written to the error queue. If there
are any errors in the error queue the front panel error annunciator is
displayed as shown in Figure 4-1.
Other errors can also be generated when the power meter is being
operated over the remote interface. These errors also display the error
annunciator and are written to the error queue.
Figure 4-1: Error Annunciator Position
To read the error queue from the front panel press:
, , then use to scroll through
each error message.
To read the error queue from the remote interface use:
the SYSTem:ERRor? command.
Error queue messages have the following format:
For example, -330, “Self-test Failed;Battery Fault”.
Error annunciator
System
Inputs
More
Error List Next
Error
,
;
Device
Error
Description
Dependent InfoNumber
HP4402.book Page 2 Thursday, November 30, 2000 1:39 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4419B and is the answer not in the manual?

Agilent Technologies E4419B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4419B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals