Power Meter Operation
Making Measurements with the Agilent E-Series Power Sensors
2-16 Agilent E4419B User’s Guide
Making Measurements with the Agilent E-Series 
Power Sensors
This section describes how to make continuous wave measurements using 
the Agilent E-series power sensors. The Agilent E-series power sensors 
have their sensor calibration tables stored in EEPROM. This allows the 
frequency and calibration data to be downloaded to the power meter 
automatically.
To make measurements, perform the following steps:
1. Zero and calibrate the power meter. 
2. Set the frequency for the signal you want to measure.
3. Take a reading.
Procedure 
The following procedure details how to make a measurement on channel A 
of the power meter. To make a measurement on channel B use the same 
procedure using the equivalent channel B softkeys. 
1. Ensure that no power is applied to the power sensor.
2. Press .
3. Press  . During zeroing, which takes approximately 10 
seconds, the wait symbol is displayed.
4. Connect the power sensor to the POWER REF output.
5. Press  ,   to calibrate the power meter. During 
calibration the wait symbol is displayed. (The power meter 
automatically turns on the POWER REF output.)
6. Press  . The current setting of the frequency is displayed 
under the   softkey. 
7. To change this setting press  . The power meter displays 
the frequency in a pop up window. Modify this frequency (see 
below) as desired.
■ Use   or   to modify the digit on which the cursor is 
currently positioned.
■ Use   or   to move to other digits.
8. To confirm your choice press the appropriate frequency units. 
9. Connect the power sensor to the signal to be measured.
Zero
Cal
Cal Cal A
A Freq
A Freq