Error Messages
Error Messages
Agilent E4419B User’s Guide 4-11
-330 Self-test Failed;Measurement Channel A Fault
Refer to “Measurement Assemblies”, on page 2-77 if you require a
description of the Measurement Assembly test.
-330 Self-test Failed;Measurement Channel B Fault
Refer to “Measurement Assemblies”, on page 2-77 if you require a
description of the Measurement Assembly test.
-330 Self-test Failed;Option 001 Battery requires replacement
The option 001 battery is not charging to a satisfactory level and should
be replaced.
-330 Self-test Failed;RAM Battery Fault
Refer to “RAM Battery”, on page 2-77 if you require a description of the
battery test.
-330 Self-test Failed;Calibrator Fault
Refer to “Calibrator”, on page 2-78 if you require a description of the
calibrator test.
-330 Self-test Failed;ROM Check Failed
Refer to “ROM Checksum”, on page 2-77 if you require a description of
the ROM Checksum test.
-330 Self-test Failed;RAM Check Failed
Refer to “RAM”, on page 2-77 if you require a description of the RAM
test.
-330 Self-test Failed;Display Assy. Fault
Refer to “Display”, on page 2-78 if you require a description of the
Display test.
-330 Self-test Failed;Confidence Check Fault ChA
Refer to “Confidence Check”, on page 2-75 if you require a description of
this test.
-330 Self-test Failed;Confidence Check Fault ChB
Refer to “Confidence Check”, on page 2-75 if you require a description of
this test.
-330 Self-test Failed;Serial Interface Fault
Refer to “Serial Interface”, on page 2-77 if you require a description of
this test.