Power Meter Operation
Zeroing and Calibrating the Power Meter
Agilent E4419B Users Guide 2-11
Press . Use the , , and hardkeys to
enter 99.4. Press .
Connect the channel A power sensor to the POWER REF output.
Press .
Remove the channel A power sensor from the POWER REF
output.
Connect the channel B power sensor to the POWER REF output.
Press .
B Ref CF
%
Cal A
Cal B