Power Meter Operation
Zeroing and Calibrating the Power Meter
Agilent E4419B User’s Guide 2-11
• Press . Use the , , and hardkeys to
enter 99.4. Press .
• Connect the channel A power sensor to the POWER REF output.
• Press .
• Remove the channel A power sensor from the POWER REF
output.
• Connect the channel B power sensor to the POWER REF output.
• Press .
B Ref CF
%
Cal A
Cal B