Power Meter Operation
Making Measurements using Sensor Calibration Tables
2-28 Agilent E4419B User’s Guide
Agilent 8481D Agilent 8485A
RCF 99 RCF 100
50 MHz 99 50 MHz 100
500 MHz 99.5 2 GHz 99.5
1GHz 99.4 4GHz 98.9
2GHz 99.5 6GHz 98.5
3GHz 98.6 8GHz 98.3
4 GHz 98.6 10 GHz 98.1
5 GHz 98.5 11 GHz 97.8
6 GHz 98.5 12 GHz 97.6
7 GHz 98.6 12.4 GHz 97.6
8 GHz 98.7 14 GHz 97.4
9 GHz 99.5 16 GHz 97
10 GHz 98.6 17 GHz 96.7
11 GHz 98.7 18 GHz 96.6
12 GHz 99 19 GHz 96
12.4 GHz 99.1 20 GHz 96.1
13 GHz 98.9 21 GHz 96.2
14 GHz 99.4 22 GHz 95.3
15 GHz 98.9 23 GHz 94.9
16 GHz 99.1 24 GHz 94.3
17 GHz 98.4 25 GHz 92.4
18 GHz 100.1 26 GHz 92.2
Agilent R8486A 26.5 GHz 92.1
RCF 100 Agilent R8486D
50 MHz 100 RCF 97.6
26.5 GHz 94.9 50 MHz 97.6
27 GHz 94.9 26.5 GHz 97.1
28 GHz 95.4 27 GHz 95.3
29 GHz 94.3 28 GHz 94.2
30 GHz 94.1 29 GHz 94.5
31 GHz 93.5 30 GHz 96.6
32 GHz 93.7 31 GHz 97.6
33 GHz 93.7 32 GHz 98
34 GHz 94.9 33 GHz 98.9
34.5 GHz 94.5 34 GHz 99.5
35 GHz 94.4 34.5 GHz 99
36 GHz 93.7 35 GHz 97.6
37 GHz 94.9 36 GHz 99
38 GHz 93.5 37 GHz 98.2
39 GHz 93.9 38 GHz 97.4
40 GHz 92.3 39 GHz 97.6
40 GHz 100
HP4402.book Page 28 Thursday, November 30, 2000 1:39 PM