Chapter 6 115
Replacement Procedure
Power Inlet Assembly Replacement
6. Replacement Procedure
Power Inlet Assembly Replacement
Tools Required
Torque screwdriver, TORX T20
Torque screwdriver, TORX T8
Torque screwdriver, TORX T10
Box torque wrench, 5/8 inch
Removal Procedure
Refer to Figure 6-14 for this procedure.
Step 1. Remove the 3/4 Rack Module Top as described in “3/4 Rack Module Top Replacement” on
page 90.
Step 2. Remove the PPMC PCA as described in “PPMC PCA Replacement” on page 101.
Step 3. Remove the A1 Mother Board as described in “A1 Mother Board Replacement” on
page 103.
Step 4. Disconnect the power cable connector (item 1) from the Power Supply Assembly.
Step 5. Remove the TORX T20 screw (item 2) fastening the ground cable.
Step 6. Remove the two TORX T10 screws (item 3) fastening the Inlet Assembly.
Step 7. Remove the Inlet Assembly (item 4) from the Rear Panel.
Step 8. Detach the fuse box (item 5) from the original Inlet Assembly and remove the fuse (item 6
).
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
NOTE When you fasten the ground cable with the TORX T20 screw, set screwdriver's torque to
21.1 in-lb.
When you fasten the Inlet Assembly with the two TORX T10 screws, set screwdriver's
torque to 8.7 in-lb.
Do not forget to insert the fuse in the new Inlet Assembly.

Table of Contents

Other manuals for Agilent Technologies E4980A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4980A and is the answer not in the manual?

Agilent Technologies E4980A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4980A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals