Chapter 4 59
Troubleshooting
Performance test failure troubleshooting
4. Troubleshooting
Performance test failure troubleshooting
Table 4-15 represents the relationships between the failed test and probable faulty
assembly. If the performance test failure cannot be removed by a proper adjustment,
replace the assembly shown in this table.
Note that this table lists some typical cases. There are possibilities that another assembly
may be faulty.
###: Most suspicious assembly
##: Suspicious assembly
#: Possible faulty assembly
Table 4-15 Performance test failure troubleshooting information
Test
Num.
First failed test Probable faulty board
assembly
A1 A2 A3 PPMC
1 Frequency Accuracy Test # ###
2 DC Source Level Accuracy Test # ###
3 DC Bias Level Accuracy Test # ###
4 DC Bias V Level Monitor Accuracy Test ###
5 DC Bias I Level Monitor Accuracy Test ###
6 AC Signal Level Monitor Accuracy Test ### #
7 AC Signal Level Accuracy Test # ###
8 Impedance Measurement Accuracy Test ### #
9 DCR Measurement Accuracy Test ### #

Table of Contents

Other manuals for Agilent Technologies E4980A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4980A and is the answer not in the manual?

Agilent Technologies E4980A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4980A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals