Chapter 6 117
Replacement Procedure
PCA E4980 USB Replacement
6. Replacement Procedure
PCA E4980 USB Replacement
Tools Required
Torque screwdriver, TORX T20
Removal Procedure
Refer to Figure 6-15 for this procedure.
Step 1. Remove the 3/4 Rack Module Top as described in “3/4 Rack Module Top Replacement” on
page 90.
Step 2. Disconnect the Front USB Cable (item 1) from PCA E4980 USB.
Step 3. Remove the two TORX T20 screws (item 2) fastening the PCA E4980 USB and the ground
cable.
Step 4. Remove the PCA E4980 USB (item 3) from the Front Panel Assembly.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
NOTE When you fasten the PCA E4980 USB and the ground cable with the two TORX T20
screws, set screwdriver's torque to 14 in-lb.
Figure 6-15 PCA E4980 USB Removal