Chapter 6 99
Replacement Procedure
Option Interface Replacement
6. Replacement Procedure
Option Interface Replacement
Tools Required
Torque screwdriver, TORX T10
Removal Procedure
Refer to Figure 6-6 for this procedure.
Step 1. Remove the two TORX T10 screws (item 1) fastening the Option Interface.
Step 2. Pulls the Option Interface carefully.
Step 3. Carefully disconnect the flat cable (item 2) of Option Interface side. Release the cable’s
connector by pressing the locking hooks at the sides of the connector.
Step 4. Remove the two TORX T10 screws (item 3) fastening the option panel.
Step 5. Remove the option panel from the option interface board.
Step 6. Disconnect the flat cable from the A1 Mother Board connector (item 4).
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
NOTE When you fasten the option panel with the two TORX T10 screws, set screwdriver's torque
to 8.7 in-lb.
When you fasten the interface card with the two TORX T10 screws, set screwdriver's
torque to 8.7 in-lb.

Table of Contents

Other manuals for Agilent Technologies E4980A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4980A and is the answer not in the manual?

Agilent Technologies E4980A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4980A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals