Appendix D 145
Error Messages
Error number: 22
D. Error Messages
error. When this error occurs, contact Agilent Technology’s Sales and Service Office or the
company from which you bought the device.
22 Fixture over temperature
The temperature inside the 42842A/B bias test fixture exceeds the limit. Turn off the DC
bias and cool down the equipment.
30 Function type not available
The Ls-Rdc, Lp-Rdc and Vdc-Idc are not available when the 42841A is connected.
G
-105 GET not allowed
GET is not allowed inside a program message.
H
70 Handler I/F not installed
The handler interface function is set to ON in a model that does not have a handler
interface board (option 201).
-114 Header Suffix out of Range
The header suffix is out of range.
I
73 I BIAS I/F not installed
The DC bias interface function is set to ON in a model that does not have the DC bias
interface board (option 002).
-224 Illegal parameter value
The parameter value is not properly set. This error occurs, for example, when the
:FUNC:IMP command is used to specify an Impedance-parameter that does not exist (as
CPRS).
51 Inconsistent limit setting
For the comparator function, user attempts to set the threshold in the sequential mode by
sending the COMP:SEQ:BIN command, although the tolerance mode is now selected.
Select the sequential mode using COMP:MODE.
-161 Invalid block data
Invalid block data was received (for example, END received before length satisfied).
-101 Invalid character
Invalid character was received.
-141 Invalid character data
Bad character data or unrecognized character data was received.
-121 Invalid character in number
Invalid character in numeric data.