702 Tuning Voltage
1. Using an oscilloscope, check 10 MHz Synthesis line on P33–15. The signal period should be
100 ns with an amplitude of approximately 3V
pp
.
If the signal is good, replace the A6 Frac–N.
2. If the signal is bad, turn the signal generator power off, remove the A6 Frac–N, and measure the
signal at P33–15 again.
If the signal is good, replace the A6 Frac–N.
3. If the signal is still bad, measure P41–5 on the A7 Reference.
If the signal is good, at P41–5, replace the A31 Motherboard.
If the signal is bad on P41–5, replace the A7 Reference.
703 Output Voltage
1. Using an oscilloscope, check 10 MHz Synthesis line on P33–15. The signal period should be
100 ns with an amplitude of approximately 3V
pp
.
If the signal is good, replace the A6 Frac–N.
2. If the signal is bad, turn the signal generator power off, remove the A6 Frac–N, and measure
P33–15 again.
If the signal is good, replace the A6 Frac–N.
3. If the signal is still bad, measure P41–5 on the A7 Reference.
If the signal is good at P41–5, replace the A31 Motherboard.
If the signal is bad at P41–5, replace the A7 Reference.

Table of Contents

Other manuals for Agilent Technologies E8267C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8267C and is the answer not in the manual?

Agilent Technologies E8267C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8267C
CategoryPortable Generator
LanguageEnglish

Related product manuals