5Replaceable Parts
108 E4416A/E4417A Power Meters Service Guide
Replacing the Front Panel Power Reference Cable Assembly
1 Remove the front panel from the power meter. (Refer to “Removing the A3
Front Panel Assembly on page 105 for details on removing the front panel).
2 Replace the power reference output cable assembly using the replacement
kit and the torques indicated on the following diagrams:
25 lb/in
6 lb/in
0 lb/in
Fold up tab after
assembly