6 Troubleshooting
118 E4416A/E4417A Power Meters Service Guide
Figure 6-1 Troubleshooting Guide - TTL Output 1
Start
Power Up
meter
POST
Pass?
Connect Sensor to Channel A
Any
Sensor
Errors?
Press
System/Inputs
More
Service
Self Test
Instrument Self Test
Self Test
Errors?
Using DVM E2373A or equivalent,
End
Yes
No
A
Select Done
No
set to VDC, monitor Rmt I/O
Connector pin 3 wrt pin 8.
Investigate Power
Meter Failure
Investigate
Sensor failure
No
Yes
Yes