Adjustments 3
E4416A/E4417A Power Meters Service Guide 71
Display Brightness and Contrast Adjustment
Introduction
The following procedure should be performed whenever a front panel
assembly or processor assembly is replaced.
The brightness is controlled automatically after executing the
softkey. Press , , and
to locate this softkey menu.
Procedure
1 Access the softkey menu and use the
softkey to increase the contrast, or the softkey to decrease the
contrast of the display.
2 When the desired display contrast level is obtained, press the
softkey, located under the softkey menu.
Set Brightness
System
More
Service
Display
NOTE
The contrast adjustment is subjective and varies according to individual user
requirements.
Display
Set Contrast
Service
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Agilent Technologies EPM-P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EPM-P Series and is the answer not in the manual?

Agilent Technologies EPM-P Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEPM-P Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals