1 Specifications
6 E4416A/E4417A Power Meters Service Guide
Average-only Mode
Absolute:Logarithmic 0.02 dB Linear 0.5%
Relative:Logarithmic 0.04 dB Linear 1.0%
Normal Mode
Absolute and relative accuracy vs. temperature is shown in the following
table:
Time Base
0.01%.
Zero Set, Zero Drift and Measurement Noise:
For E- Series E9320 sensors, refer to the Specifications Section in the E- Series
E9320 Power Sensors Operating and Service Guide.
Calibration Temperature
1
±5 ºC
Temperature 0 to 55 ºC
Absolute accuracy (log) ±0.04 dB ±0.08 dB
Absolute accuracy (linear) ±0.8% ±1.7%
Relative accuracy (log) ±0.08 dB ±0.16 dB
Relative accuracy (linear) ±1.6% ±3.4%
1. Power meter is within 5 ºC of its calibration temperature.