Measurement Uncertainty Analysis - Power Reference Level Test C
E4416A/E4417A Power Meters Service Guide C-141
Mismatch
The Mismatch uncertainty was previously determined as being:
Consider the model below:
An estimate of the moduli of P
d
and P
s
are each less than a specified value. P
s
and P
d
each lie within a circle of radius P. Assuming P
d
and P
s
have equal
probability of lying anywhere within the circle the standard uncertainty of
Mismatch becomes
P
s
is the reflection coefficient of the source, which is specified as less than
1.05.
VSWR < 1.05
M
s
Unc erta inty mag nitude()2P
s
P
d
=
NOTE
This is the worst-case uncertainty for mismatch and occurs when the incident and
reflected waves are 180 degrees out of phase
P
uM
s
()
maxP
s
maxP
d
2
-----------------------------------------
±=
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Agilent Technologies EPM-P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EPM-P Series and is the answer not in the manual?

Agilent Technologies EPM-P Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEPM-P Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals