Chapter 3 3-39
Functional Testing
Resolution Bandwidth Switching Uncertainty
5. Repeat step 3 and 4 for each of the remaining resolution bandwidth
and span settings listed in Table 3-6. The Mkr 1 amplitude reading
should be within the range indicated in the table “Test Limits” on
page 3-37.
Table 3-6 Test Limits Table
Resolution
Bandwidth Setting
SPAN Setting Mkr 1 Amplitude Reading
1 kHz 50 kHz 0 (Ref)
3 kHz 50 kHz
10 kHz 50 kHz
30 kHz 500 kHz
100 kHz 500 kHz
300 kHz 5 MHz
1 MHz 10 MHz
3 MHz 10 MHz
5 MHz 50 MHz