120 5977 Series MSD Operation Manual
5 General Maintenance
Before Starting
You can perform much of the maintenance required by your MSD. For your
safety, read all of the information in this introduction before performing any
maintenance tasks.
Scheduled maintenance
Common maintenance tasks are listed in Table 21. Performing these tasks
when scheduled can reduce operating problems, prolong system life, and
reduce overall operating costs.
Table 21 Maintenance schedule
Task Every week Every 6 months Every year As needed
Tune the MSD X
Check the foreline pump oil level X
Check the calibration vial(s) X
Replace the foreline pump oil
*
X
Replace the diffusion pump fluid X
Check the dry foreline pump X
Change the dry foreline pump tip seal X
Change the foreline pump exhaust filter X
Clean the ion source X
Check the carrier gas trap(s) on the GC and MSD X
Replace the worn out parts X
Lubricate sideplate or vent valve O-rings
X
Replace CI Reagent gas supply X
Replace GC gas supplies X
* Every 3 months for CI MSDs using ammonia reagent gas.
Vacuum seals other than the side plate O-ring and vent valve O-ring do not need to be lubricated. Lubricating other seals
can interfere with their correct function.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7035A and is the answer not in the manual?

Agilent Technologies G7035A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7035A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals