Operating in Electron Ionization (EI) Mode 3
5977 Series MSD Operation Manual 49
<method> Complete <timestamp>
The run and subsequent data processing are done. The same message appears
even if the run was terminated prematurely.
Method Loaded <method name>
Method parameters were sent to the MSD.
MS Locked by <computer name>
MS parameters can only be changed from the MassHunter Data Acquisition.
Press Sideplate
A reminder during startup to press the MSD side plate to ensure an adequate
vacuum seal.
Run: <method> Acquiring <datafile>
A run is in progress; data is being acquired to the designated data file.
To view system status during startup
1 The following messages are displayed on the LCP display during startup:
Press Sideplate
Loading OS
Press Sideplate
Loading MSD Firmware
2 Continue to press the sideplate of the MSD until the MSD Ready message
appears. This helps the instrument to pump down more quickly.