34 5977 Series MSD Operation Manual
2 Installing GC Columns
To Install a Capillary Column in a Split/Splitless Inlet
Materials needed
Gloves, clean
Large (8650-0030)
Small (8650-0029)
Metric ruler
Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)
Capillary column
Column cutter, ceramic (5181-8836) or diamond (5183-4620)
Ferrules
0.27-mm id, for 0.10-mm id columns (5062-3518)
0.37-mm id, for 0.20-mm id columns (5062-3516)
0.40-mm id, for 0.25-mm id columns (5181-3323)
0.5-mm id, for 0.32-mm id columns (5062-3514)
0.8-mm id, for 0.53-mm id columns (5062-3512)
Inlet column nut (5181-8830 for Agilent 7890 Series and 7820)
Magnifying loupe
Septum (may be old, used inlet septum)
To install columns in other types of inlets, refer to your Gas Chromatograph
User Information.
WARNING
The GC operates at high temperatures. In order to avoid burns, do not touch any
parts of the GC until you are sure they are cool.
WARNING
Always wear safety glasses when handling capillary columns. Use care to avoid
puncturing your skin with the end of the column.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7035A and is the answer not in the manual?

Agilent Technologies G7035A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7035A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals