Installing GC Columns 2
5977 Series MSD Operation Manual 33
Conditioning columns
Conditioning a column before it is connected to the GC/MSD interface is
essential. See “To Condition a Capillary Column” on page 37.
A small portion of the capillary column stationary phase is often carried away
by the carrier gas. This is called column bleed. Column bleed deposits traces of
the stationary phase in the MSD ion source. This decreases MSD sensitivity
and makes cleaning the ion source necessary.
Column bleed is most common in new or poorly crosslinked columns. It is
much worse if there are traces of oxygen in the carrier gas when the column is
heated. To minimize column bleed, all capillary columns should be
conditioned before they are installed in the GC/MSD interface.
Conditioning ferrules
Heating ferrules to their maximum expected operating temperature a few
times before they are installed can reduce chemical bleed from the ferrules.
Tips and hints
The column installation procedures for the 5977 Series MSDs may be
different from that for previous MSDs. Using the procedure from another
instrument may not work and may damage the column or the MSD.
You can remove old ferrules from column nuts with an ordinary push pin.
Always use carrier gas that is at least 99.9995% pure.
Because of thermal expansion, new ferrules may loosen after heating and
cooling a few times. Check for tightness after two or three heating cycles.
Always wear clean gloves when handling columns, especially the end that
will be inserted into the GC/MSD interface.
WARNING
If you are using hydrogen as a carrier gas, do not start carrier gas flow until the
column is installed in the MSD and the MSD has been pumped down. If the vacuum
pumps are off, hydrogen will accumulate in the MSD and an explosion may occur.
See “Hydrogen Safety” on page 19.
WARNING
Always wear safety glasses when handling capillary columns. Use care to avoid
puncturing your skin with the end of the column.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7035A and is the answer not in the manual?

Agilent Technologies G7035A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7035A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals