162 5977 Series MSD Operation Manual
6 CI Maintenance
To Clean the CI Ion Source
Materials needed
Abrasive paper (5061-5896)
Alumina abrasive powder (8660-0791)
Aluminum foil, clean
Cloths, clean (05980-60051)
Cotton swabs (5080-5400)
Glass beakers, 500 mL
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Solvents
Acetone, reagent grade
Methanol, reagent grade
Methylene chloride, reagent grade
Ultrasonic bath
Preparation
1 Disassemble the ion source. See “To Disassemble the CI Ion Source” on
page 159.
2 Collect the following parts to be cleaned for a CI ion source: (Figure 30 on
page 163)
Repeller
Source body
Drawout plate
Drawout cylinder
Ion focus lens
Entrance lens
These are the parts that contact the sample or ion beam. The other parts
normally should not require cleaning.