General Maintenance 5
5977 Series MSD Operation Manual 123
Chemical residue
Only a small portion of your sample is ionized by the ion source. The majority
of any sample passes through the ion source without being ionized. It is
pumped away by the vacuum system. As a result, the exhaust from the foreline
pump will contain traces of the carrier gas and your samples. Exhaust from
the standard foreline pump also contains tiny droplets of foreline pump oil.
An oil trap is supplied with the standard foreline pump. This trap stops only
pump oil droplets. It does not trap any other chemicals. If you are using toxic
solvents or analyzing toxic chemicals, do not use this oil trap. For all foreline
pumps, install a hose to take the exhaust from the foreline pump outdoors or
into a fume hood vented to the outdoors. For the standard foreline pump, this
requires removing the oil trap. Be sure to comply with your local air quality
regulations.
The fluids in the diffusion pump and standard foreline pump also collect
traces of the samples being analyzed. All used pump fluid should be
considered hazardous and handled accordingly. Dispose of used fluid
correctly, as specified by your local regulations.
Electrostatic discharge
All of the printed circuit boards in the MSD contain components that can be
damaged by electrostatic discharge (ESD). Do not handle or touch these
boards unless absolutely necessary. In addition, wires, contacts, and cables
can conduct ESD to the electronics boards to which they are connected. This is
especially true of the mass filter (quadrupole) contact wires which can carry
WARNING
The oil trap supplied with the standard foreline pump stops only foreline pump oil. It
does not trap or filter out toxic chemicals. If you are using toxic solvents or
analyzing toxic chemicals, remove the oil trap. Do not use the trap if you have a CI
MSD. Install a hose to take the foreline pump exhaust outside or to a fume hood.
WARNING
When replacing pump fluid, use appropriate chemical-resistant gloves and safety
glasses. Avoid all contact with the fluid.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7035A and is the answer not in the manual?

Agilent Technologies G7035A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7035A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals