CI Maintenance 6
5977 Series MSD Operation Manual 159
To Disassemble the CI Ion Source
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Hex ball driver, 1.5 mm (8710-1570)
Hex ball driver, 2.0 mm (8710-1804)
Wrench, open-end, 10 mm (8710-2353)
Procedure
1 Remove the ion source. See “To Remove the CI Ion Source” on page 157.
2 Remove the filaments. Refer to Figure 29 on page 160.
3 Separate the repeller assembly from the source body. The repeller
assembly includes the source heater block assembly, repeller, and related
parts.
4 Remove the repeller and ceramic insulator and separate them.
5 Remove the setscrew for the lenses.
6 Pull the lens assembly out of the source body.
7 Remove the drawout cylinder and drawout plate from the source body.
8 Separate the ion focus lens, entrance lens and insulator.