Agilent InfinityLab LC Series VWD User Manual 107
8Test Functions
Intensity Test
Intensity Test Failed
Intensity Test Failed
Probable cause Suggested actions
1
Empty flow cell
Ensure the flow cell is filled with water.
2
Flow cell windows dirty
Repeat the test with the flow cell window
removed. If the test passes, exchange the flow
cell windows.
3
Optics defect
Please contact your Agilent service representa-
tive.
4
Defective lamp or optics.
Exchange the lamp.