Agilent InfinityLab LC Series VWD User Manual 93
7Error Information
Detector Error Messages
Wavelength holmium check failed
Error ID: 7318
The holmium oxide test in the detector has failed. During the holmium test, the
detector moves the holmium filter into the light path, and compares the
measured absorbance maxima of the holmium oxide filter with expected
maxima. If the measured maxima are outside the limits, the error message is
generated.
Probable cause Suggested actions
1
Misaligned/defective grating assembly.
Ensure the flow cell is inserted correctly, and
is free from contamination (cell windows,
buffers, and so on).
Run the filter-motor test to determine if the
filter motor assembly is defective. If defec-
tive, please contact your Agilent service rep-
resentative.
Run the grating-motor test to determine if
the grating assembly is defective. If defec-
tive, please contact your Agilent service rep-
resentative.

Table of Contents

Other manuals for Agilent Technologies G7114A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7114A and is the answer not in the manual?

Agilent Technologies G7114A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7114A
CategorySecurity Sensors
LanguageEnglish

Related product manuals