Agilent InfinityLab LC Series VWD User Manual 118
8Test Functions
Holmium Oxide Test
Interpreting the Results
The test is passed successfully when all three wavelengths are within ± 1 nm of
the expected value. This indicates the detector is calibrated correctly.
1 Run the Holmium Oxide Test with the Agilent Lab Advisor (for further
information see Online- Help of user interface).
Figure 36 Holmium Oxide Test - Results
Figure 37 Holmium Oxide Test - Signals