Agilent InfinityLab LC Series VWD User Manual 92
7Error Information
Detector Error Messages
Wavelength calibration setting failed
Error ID: 7310
The intensity maximum was not found during wavelength calibration.
Calibration 0 Failed: Zero-order calibration failed.
Calibration 1 Failed: 656 nm calibration failed.
Probable cause Suggested actions
1
Lamp is OFF.
Switch on the lamp.
2
Incorrect flow cell installation.
Ensure the flow cell is installed correctly.
3
Flow cell contamination or air bubbles.
Clean/replace flow cell windows or remove air
bubbles.
4
Intensity too low.
Replace lamp.
5
Current step value too far from maximum.
Repeat the calibration.
Please contact your Agilent service represen-
tative.
6
Misaligned/defective grating assembly.
Please contact your Agilent service representa-
tive.
7
Defective main board.
Please contact your Agilent service representa-
tive.

Table of Contents

Other manuals for Agilent Technologies G7114A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7114A and is the answer not in the manual?

Agilent Technologies G7114A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7114A
CategorySecurity Sensors
LanguageEnglish

Related product manuals