Agilent InfinityLab LC Series VWD User Manual 146
9Maintenance
Replace the Module Firmware
Replace the Module Firmware
To upgrade/downgrade the module’s firmware carry out the following steps:
1 Download the required module firmware, the latest FW Update Tool and the
documentation from the Agilent web.
http://www.agilent.com/en-us/firmwareDownload?whid=69761
2 For loading the firmware into the module follow the instructions in the
documentation.
Module Specific Information
When
The installation of newer firmware might be necessary
• if a newer version solves problems of older versions or
• to keep all systems on the same (validated) revision.
The installation of older firmware might be necessary
• to keep all systems on the same (validated) revision or
• if a new module with newer firmware is added to a system or
• if third party control software requires a special version.
Tools required Description
Agilent Lab Advisor software
Parts required # Description
1 Firmware, tools and documentation from Agilent web site
Preparations
Read update documentation provided with the Firmware Update Tool.
Table 14 Module Specific Information (G7114A/B)
G7114B G7114A
Initial firmware D.06.70 D.07.01
Compatibility with 1100 / 1200
series modules
When using the G7114A/B in a system, all other modules must
have firmware from set 6.50 or above (main and resident). Other-
wise the communication will not work.
Conversion / emulation Possible via Lab Advisor software:
• G7114B: G1314E/F
• G7114A: G1314F