Agilent InfinityLab LC Series VWD User Manual 29
2 Site Requirements and Specifications
Physical Specifications
Physical Specifications
Table 2 Physical Specifications
Type Specification Comments
Weight 11 kg (24.3 lbs)
Dimensions
(height × width × depth)
140 x 396 x 436 mm
(5.5 x 15.6 x 17.2 inches)
Line voltage 100 – 240 V~, ± 10 % Wide-ranging capability
Line frequency 50 or 60 Hz, ± 5 %
Power consumption 80 VA, 70 W
Ambient operating tempera-
ture
4 - 55 °C (39 - 131 °F)
Ambient non-operating tem-
perature
-40 – 70 °C (-40 – 158 °F)
Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Non-operating altitude Up to 4600 m (15092 ft) For storing the module
Safety standards:
IEC, EN, CSA, UL
Installation category II, Pollution degree 2 For indoor use only.
ISM Classification ISM Group 1 Class B According to CISPR 11