Agilent InfinityLab LC Series VWD User Manual 109
8Test Functions
Cell Test
Cell Test Failed
Cell Test Failed
Probable cause Suggested actions
1
Cell contaminated
Flush flow cell
2
Cell windows are contaminated
Clean/replace cell windows
3
Mechanical problem
Check cell position