EasyManuals Logo
Home>Agilent Technologies>Security Sensors>G7114A

Agilent Technologies G7114A User Manual

Agilent Technologies G7114A
222 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #185 background imageLoading...
Page #185 background image
Agilent InfinityLab LC Series VWD User Manual 185
12 Hardware Information
Interfaces
5V Distribution (Future Use)
Available directly after turn on oft the hosting module (assures that certain
base functionality of the device can be detected by firmware).
For digital circuits or similar.
Provided 500 mA maximum.
Short-circuit proof with automatic switch off (by firmware).
24V Distribution (Future Use)
Available by firmware command (defined turn on/off).
For devices that need higher power
Class 0: 0.5 A maximum (12 W)
Class 1: 1.0 A maximum (24 W)
Class 2: 2.0 A maximum (48 W)
Class depends on hosting module’s internal power overhead.
If a connected device requires more power the firmware detects this
(overcurrent detection) and provides the information to the user interface.
Fuse used for safety protection (on board).
Short circuit will be detected through hardware.
USB (Universal Serial Bus)
USB (Universal Serial Bus) - replaces RS232, supports:
a PC with control software (for example Agilent Lab Advisor)
USB Flash Disk

Table of Contents

Other manuals for Agilent Technologies G7114A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7114A and is the answer not in the manual?

Agilent Technologies G7114A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7114A
CategorySecurity Sensors
LanguageEnglish

Related product manuals