Agilent InfinityLab LC Series VWD User Manual 121
8Test Functions
D/A Converter (DAC) Test
Running the test with Agilent Lab Advisor
1 Run the D/A Converter Test with the Agilent Lab Advisor (for further
information see Online- Help of user interface).
Figure 38 Converter Test - Results
Figure 39 D/A Converter (DAC) Test – Example of Integrator Plot
Running the Test with Instant Pilot
The test can be started via the command line.
1 To start the test
TEST: DAC 1
Reply: RA 00000 TEST:DAC 1
2 To stop the test
TEST:DAC 0
Reply: RA 00000 TEST:DAC 0