Agilent InfinityLab LC Series Vialsamplers User Manual 134
3Using the Module
Transporting the Sampler
1 Turn off the Sample Cooler/Sample Thermostat.
2 Remove condensate.
a Place a suitable container underneath the outlet pipe.
b Remove the drainage tube.
c Gently tap the sides of the sampler several times to facilitate the drainage
of the condensate from the system.
3 Remove all vials from the vial drawers.
4 Move the transport assembly to the park position using Instant Pilot or Lab
Advisor, see
“Park Arm on page 186.
5 Turn off the sampler.
6 Install the Transport Protection Foam, see “Install the Transport Protection
Foam” on page 135.
7 Remove the Sample Cooler/Sample Thermostat from the sampler if needed,
see
“Replace the Sample Cooler/Sample Thermostat” on page 270.
NOTE
Do not tilt the module to avoid damage to the internal parts.