EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>G7129A

Agilent Technologies G7129A User Manual

Agilent Technologies G7129A
369 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #194 background imageLoading...
Page #194 background image
Agilent InfinityLab LC Series Vialsamplers User Manual 194
7Error Information
General Error Messages
Shutdown
Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status signals.
A LOW signal input on pin 4 of the remote connector generates the error
message.
Probable cause Suggested actions
1
Leak detected in another module with a CAN
connection to the system.
Fix the leak in the external instrument before
restarting the module.
2
Leak detected in an external instrument with
a remote connection to the system.
Fix the leak in the external instrument before
restarting the module.
3
Shut-down in an external instrument with a
remote connection to the system.
Check external instruments for a shut-down con-
dition.
4
The degasser failed to generate sufficient
vacuum for solvent degassing.
Check the vacuum degasser for an error condi-
tion. Refer to the Service Manual for the degasser
or the pump that has the degasser built-in.

Table of Contents

Other manuals for Agilent Technologies G7129A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7129A and is the answer not in the manual?

Agilent Technologies G7129A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7129A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals