Agilent InfinityLab LC Series Vialsamplers User Manual 270
8Maintenance
Replace the Sample Cooler/Sample Thermostat
Replace the Sample Cooler/Sample Thermostat
For the Sample Cooler installation in a Vialsampler the serial number of the
Sample Cooler must be DEBAT02001 or higher.
When
If the Sample Cooler/Sample Thermostat is damaged or defective.
Tools required p/n Description
8710-0899 Screwdriver Pozidrive Shaft
(for the Sample Cooler)
5182-3466 Torx screwdriver T10
(for the Sample Thermostat)
OR 5023-3089 Torx key set
Parts required # p/n Description
1 G7167-60005 Sample Cooler
OR 1 G7167-60101 Sample Thermostat
Flammable refrigerant
Formation of flammable gas-air mixtures inside the Sample Thermostat
and laboratory.
Keep open fire or sources of ignition away from the device.
Ensure a room size of 4 m
3
(1 m
3
for every 8 g of R600a refrigerant
inside of the Sample Thermostat).
Ensure adequate ventilation: typical air exchange of 25 m
3
/h per m
2
of
laboratory floor area.
Keep all ventilation openings in the enclosure clear of obstructions. Do
not block the openings on the circumference of the Sample Thermostat.
Flammable refrigerant used
When handling, installing and operating the Sample Thermostat, care
should be taken to avoid damage to the refrigerant tubing or any part of
the Sample Thermostat.