Agilent InfinityLab LC Series Vialsamplers User Manual 210
7Error Information
Vialsampler Error Messages
Cannot move metering, the metering type definition
is incorrect
Error ID: 34320
The analytical head is not operational because there is no valid information
available for its type in the firmware.
Initialization failed errors
These are non-specific error messages and indicate only that something went
wrong during the initialization of the given component. They are usually coupled
with other error events, which provide more specific information on the nature of
the issue. To identify these source error events, look for entries with the same
timestamp as for the Initialization failed error message.
• Error ID: 34322 → Metering device initialization failed
• Error ID: 34323 → Injection valve initialization failed
• Error ID: 34324 → Needle device initialization failed
• Error ID: 34325 → Transport initialization failed
• Error ID: 34326 → Sampler initialization failed
Probable cause Suggested actions
1
Missing configuration data
Verify that the type of the analytical head is cor-
rectly configured in the CDS or Lab Advisor.
2
Potential hardware error
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1
Initialization failed
Identify the source error message and follow the
instructions provided for that.