Agilent InfinityLab LC Series Vialsamplers User Manual 244
8Maintenance
Exchange the Sample Loop Assembly
9 Click on Next in the LabAdvisor to lift the Needle up
again.
OR
In the Local Controller, move the needle up to the upper-
most position.
10 Move the wash port back into its normal position by
turning it to the right.
11 Install the safety cover (1.) and fix the screw (2.). 12 In Lab Advisor, select Next to move the needle into the
needle seat and then Back to finish the maintenance
procedure.
OR
In the Local Controller, select Continue to move the nee-
dle into the needle seat and then Exit to finish the main-
tenance procedure.
13 Close the doors. 14 Perform a System Pressure Test.