Technical Information
Vacuum Seals
5
MS40+ Single Stage Rotary Vane Pumps User Manual / 87-901-008-01 (H.00) 63/104
Th
e pump is equipped with a special anti-suckback device which
automatically isolates the vacuum system when the pump stops. This
avoids rises in pressure or oil flow in the vacuum system.
Vacuum Seals
A special feature of this pump is the low number of gaskets that are
employed.
The seals in the circuit are obtained by means of FKM gaskets.
Sealing of the rotor shaft is guaranteed by a rotating gasket with
dust-guard lip.
The suction flange and duct are sealed by mean of OR gaskets.
Anti-Suckback Device
The pump is equipped with a special anti-suckback device to avoid
air pressure rises and/or oil back-flow towards the evacuated
chamber when the pump is switched off. This device has a shutter
which automatically closes the suction duct.
In this way the pump and vacuum system are completely isolated
from each other and air can enter the pump without any risk for the
vacuum produced in the system.
The device includes some special features, namely:
drive obtained avoiding any form of contamination of the inlet
duct by fluids (oil and/or air). Thanks to this, when the pump is
started again, the pumpdown to vacuum conditions is extremely
fast as these contaminants are not present and no degassing is
therefore required.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies MS40+ and is the answer not in the manual?

Agilent Technologies MS40+ Specifications

General IconGeneral
BrandAgilent Technologies
ModelMS40+
CategoryWater Pump
LanguageEnglish

Related product manuals