RS 232/RS 485 Communication Description
5
MS40+ Single Stage Rotary Vane Pumps User Manual / 87-901-008-01 (H.00) 91/104
So
urce: Pump
Destination: PC
02 80 06
03
38
35
STX ADDR ACK ETX CRC
Command: READ PUMP STATUS
Source: PC
Destination: Pump (with address = 3)
02
83
32
30
35
30
03
38
37
STX
ADDR
WINDOW
RD
ETX
CRC
Source: Pump (with address = 3 in stop status)
Destination: PC
02
83
32
30
35
30
30
30
30
30
30
03
38
37
STX
ADDR
WINDOW
DATA (STATUS)
ETX
CRC
Window Meanings
Tab. 9
WIN TYPE R/W DESCRIPTION
000
L
RW
START/STOP (1= START ; 0= STOP)
008 L RW REMOTE/SERIAL CONFIGURATION (1= Remote ; 0= Serial)
102 N RW SET POINT VALUE [Hz] (0 100)
105 N RW SET POINT HYSTERESIS [%] (0 100)
108 N RW BAUD RATE (0-4) [600, 1200, 2400, 4200, 9600]
120 N RW ROTATIONAL FREQUENCY SETTING [Hz] (40 - 60)
127 N R ROTATIONAL FREQUENCY SETTING [rpm] (read only)
200
N
R
BUS CURRENT [mA]
201 N R 3PHASE VOLTAGE [Vrms]
202 N R POWER [W]
203 N R DRIVING FREQUENCY [Hz]
205 N R STATUS (0=stop; 1=wait interlock; 2=start; 3=autotuning; 5=normal; 6=fail)